Maintenance of Prophy-Mate neo
• Cleaning of Prophy-Mate neo after use and regular maintenance are required.
( Failure to do so may cause powder blockage, water leakage, etc. )

• If cleaning powder remains in the powder case, powder may solidify, and may cause clogging

• Before use, thoroughly dry the coupling, handpiece, hose joint section of the powder case, inside of the powder case, and nozzle. Also, perform idle running with air alone to blow off any internal water.

• In order to prevent clogging with powder, make sure to perform ultrasonic cleaning once a week.

 

 
Troubleshooting - Powder blockage

Remove the caked powder with the nozzle/cleaning wire.


Applying too much force may bend the wire.

   
   

Perform ultrasonic cleaning and autoclave sterilization after removing the caked powder.

 

 

 

 
Troubleshooting - Water leakage

Replace the O-rings at the handpiece joint section of the powder case. Apply a drop of the provided turbine oil and rub it with a finger or other materials.