Maintenance of VarioSurg |
|
|
![]() |
Exchanging the Halogen Lamp
Remove the rear cover from the handpiece
only when replacing the lamp. Turn the rear cover in the arrow direction and
remove it. Prior to lamp replacement, disconnect the handpiece
out from the control unit. |
![]() |
Turn the rear cover and slide the cord backward. |
![]() |
Slowly pull the handpiece out. |
|
Slowly pull the ring out. |
|
| CInsert a thin slotted screwdriver etc. into
the highlighted area and remove the lamp. Securely insert a
connecting terminal of a new lamp to the socket hole. Place the rear cover and
secure it to the handpiece by screwing clockwise. |
|