Exchanging the Halogen Lamp

 

Remove the rear cover from the handpiece only when replacing the lamp. Turn the rear cover in the arrow direction and remove it. Prior to lamp replacement, disconnect the handpiece out from the control unit.

Turn the rear cover and slide the cord backward.

Slowly pull the handpiece out.

Slowly pull the ring out.

CInsert a thin slotted screwdriver etc. into the highlighted area and remove the lamp. Securely insert a connecting terminal of a new lamp to the socket hole. Place the rear cover and secure it to the handpiece by screwing clockwise.